HD Ultra X Service Guide
11. Select camera A:
12. Move the motor backwards and keep it in backwards position.
13. Turn the Vertical Position Adjustment Wheel (8) until the 1-8 lines are seen at either end.
14. Loosen the Camera Unit Fixing Screws (5).
15. Adjust the Tilt Adjustment Screw until start and end have same number of lines. If necessary
also use Vertical Position Adjustment Wheel (8) Tighten the Camera Unit Fixing Screws (5).
16. Repeat for other cameras (select B, C, D, E in turn).
17. Select CAT.
18. Select Stepper Motor Cont. RUN. This will centre the scan lines on the glass plate.
19. Select Exit to return to software oscilloscope screen.