Lid Sensor Adjustment
Tools:
After refitting the OTAC, or if the Lid always shows as open, use these instructions to make adjustments to
the sensor.
1. Press down on the Guide Plate until both ends lock in place to close.
2. With the scanner powered up, and the OTAC pushed down, with the added +1 click on the
ratchet, make sure the LED for the raised lid is off.
T10 Torx Screwdriver
HD Ultra X Service Guide
Page 101