Transcutaneous Pacemaker Analyzer Specifications (Impulse 7000Dp Only) - Fluke Biomedical Impulse 6000D User Manual

Defibrillator analyzer, defibrillator/transcutaneous pacer analyzer
Hide thumbs Also See for Impulse 6000D:
Table of Contents

Advertisement

Impulse 6000D, 7000DP
Users Manual

Transcutaneous Pacemaker Analyzer Specifications (Impulse 7000DP only)

Test Load Selections
Defibrillator Input
Fixed Load ..........................................................50 Ω
Accuracy .............................................................±1 %, non-inductive (<2 μH)
Power Rating.......................................................10 defib pulses of 360 J every 5 minutes
Pacemaker Input
Variable Load ......................................................50 to 1500 Ω in 50 Ω steps
Accuracy .............................................................±1 %, non-inductive (<2 μH)
Power Rating.......................................................5 W (average), 40 W (peak) @ 1000 Ω
Measurements
Manufacturer Specific Algorithms (plus general purpose default algorithm selection)
GE Responder (1500 & 1700)
MDE 300 (Medical Data Electronics)
Medtronic ERS/Physio Control LIFEPAK
MRL (Medical Research Laboratory/Welch Allyn)
Philips/Agilent/HP
Schiller Medical
ZOLL Medical
(plus a general purpose default algorithm selection)
Current
Range..................................................................4.00 to 250 mA
Accuracy .............................................................±(1% of reading + 0.02 mA)