Fluke Biomedical ESA614 Getting Started Manual

Electrical safety analyzer

Advertisement

Quick Links

ESA614
Electrical Safety Analyzer
Getting Started Manual
PN 5006602
August 2018, Rev. 1
© 2018 Fluke Corporation. All rights reserved. Specifications are subject to change without notice.
All product names are trademarks of their respective companies.

Advertisement

Table of Contents
loading

Summary of Contents for Fluke Biomedical ESA614

  • Page 1 ESA614 Electrical Safety Analyzer Getting Started Manual PN 5006602 August 2018, Rev. 1 © 2018 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their respective companies.
  • Page 2 Fluke Biomedical warrants this instrument against defects in materials and workmanship for one year from the date of original purchase OR two years if at the end of your first year you send the instrument to a Fluke Biomedical service center for calibration.
  • Page 3 Copyright Release Fluke Biomedical agrees to a limited copyright release that allows you to reproduce manuals and other printed materials for use in service training programs and other technical publications. If you would like other reproductions or distributions, submit a written request to Fluke Biomedical.
  • Page 4 ServiceDesk@fluke.com To ensure the accuracy of the Product is maintained at a high level, Fluke Biomedical recommends the product be calibrated at least once every 12 months. Calibration must be done by qualified personnel. Contact your local Fluke Biomedical representative for calibration.
  • Page 5: Table Of Contents

    Table of Contents Title Page Introduction ........................1 Intended Use ........................3 Safety Information ......................3 Unpack the Product ......................5 Instrument Familiarization ....................6 How to Hold the Product ....................10 Connect to Line Power ....................10 Connect a DUT to the Product ..................11 Turn On the Product .......................
  • Page 6 ESA614 Getting Started Manual Replaceable Parts ......................16 Accessories ........................17 Specifications ........................ 18 Detailed Specifications ....................20...
  • Page 7: Introduction

    Product. • Insulation resistance • Ground leakage The Fluke Biomedical ESA614 Electrical Safety Analyzer (the Product) is a full-featured, compact, portable • Chassis leakage analyzer, designed to verify the electrical safety of medical •...
  • Page 8 ESA614 Getting Started Manual • Point to point leakage, voltage, and resistance • ECG simulation and performance waveforms Symbol Description Table 1 is a list of the symbols used on the Product and in Conforms to European Union directives. this manual.
  • Page 9: Intended Use

    Electrical Safety Analyzer Intended Use Safety Information Intended Use In this manual, a Warning identifies conditions and The Product is an electronic signal source and procedures that are dangerous to the user. A Caution measurement device for verifying the electrical safety of identifies conditions and procedures that can cause medical devices.
  • Page 10 ESA614 Getting Started Manual • • Do not touch voltages >30 V ac rms, 42 V Use only current probes, test leads, and ac peak, or 60 V dc. adapters supplied with the Product. • • Do not use the Product around explosive...
  • Page 11: Unpack The Product

    • Remove all probes, test leads, and have these items: accessories that are not necessary for the measurement. • ESA614 • Disable the Product if it is damaged. • Getting Started Manual • Do not use the Product if it is damaged.
  • Page 12: Instrument Familiarization

    ESA614 Getting Started Manual Instrument Familiarization Figure 1 and Table 0-2 show the front-panel controls and connections of the Product. gtv116.eps Figure 1. Front-Panel Controls and Connections...
  • Page 13 Electrical Safety Analyzer Instrument Familiarization Table 2. Top-Panel Controls and Connections Item Name Description Equipment Outlet Controls the configuration of the equipment outlet. Opens and closes the neutral and  Configuration Buttons ground connection and reverses the polarity of the neutral and hot connection. Illuminates when high voltage is applied to the ECG/Applied Parts posts RED V/e/A High Voltage Indicator ...
  • Page 14 ESA614 Getting Started Manual Figure 2 and Table 3 describe the side and top-panel connections of the Product. gtv110.eps Figure 2. Side and Top-Panel Connections...
  • Page 15 Electrical Safety Analyzer Instrument Familiarization Table 3. Side and Top-Panel Connections Item Name Description Equipment outlet, specified to the version of the Product, which supplies a DUT Equipment Outlet  connection.  USB A Controller Port For external keyboard, barcode reader, or printer. USB Device Port Digital connection to control the Product from a PC or instrument controller.
  • Page 16: How To Hold The Product

    ESA614 Getting Started Manual Connect to Line Power How to Hold the Product When you move the Product, use the handle on the XWWarning bottom case to hold it. See Figure 3. To prevent possible electrical shock, fire, or personal injury: •...
  • Page 17: Connect A Dut To The Product

    Electrical Safety Analyzer Connect a DUT to the Product The Product is intended for use with single-phase, Turn On the Product grounded power. It is not intended for dual, split-phase or Note three-phase power configurations. It can be used with a power system that supplies the correct voltages for To make sure the high-voltage indicator works, single-phase and is grounded, or is an isolated power...
  • Page 18: How To Set The Language

    ESA614 Getting Started Manual How to Set the Language The Product can show data in English, French, German, Spanish, Italian, or Portuguese. To change the language: Push . From the Setup menu, push the Instrument Setup softkey. Push  or  until the Language variable is highlighted.
  • Page 19 Table 4. DUT Connections to the Product gtv113.eps Item Description Connect the ESA614 to grounded mains socket.  To protective earth or any exposed conductive surface on the enclosure.  Connect the DUT ac power cord to the equipment outlet on the Analyzer.
  • Page 20: For More Information

    Use only specified replacement fuses. • Have an approved technician repair the For more information on how to use the Product, refer to the ESA614 Users Manual at www.flukebiomedical.com. Product. Maintenance The Product is a calibrated measurement instrument. Use the necessary precautions to prevent mechanical abuse XW Warning that could change the calibrated adjustments.
  • Page 21: How To Clean The Product

    Applicable fuse ratings are posted on the case bottom label of the Product. Table 5 is a list of available fuses with Fluke Biomedical part numbers. Reinstall the fuse door and secure it with the screw.
  • Page 22: Replaceable Parts

    Getting Started Manual Replaceable Parts Table 5 is a list of replaceable parts for the Product. Table 5. Replaceable Parts Item Fluke Biomedical Part Number ESA614 Getting Started Manual 5006602 15 A Power Cord 2238644 Null Post Adapter 3326842 Ansur, CD with demo version...
  • Page 23: Accessories

    Electrical Safety Analyzer Accessories Accessories Table 6 is a list of available accessories for the Product. Table 6. Accessories Item Fluke Biomedical Part Number Test Leads with Retractable Sheath 1903307 Ground Pin Adapters 2242165 1-to-10 ECG Adapter 3392119 Universal Snap to Banana Adapter (10/pack)
  • Page 24: Specifications

    ESA614 Getting Started Manual Specifications Temperature Operating ............10 °C to 40 °C (50 °F to 104 °F) Storage ............... -20 °C to 60 °C (-4 °F to 140 °F) Humidity ..............10 % to 90 % non-condensing Altitude 100 V/115 V ac mains supply voltage ....5000 m 230 V ac mains supply voltage ......
  • Page 25 Electrical Safety Analyzer Specifications Safety General..............IEC 61010-1: Overvoltage Category II, Pollution Degree 2 Measurement ............IEC 61010-2-030: CAT II 300 V IP Rating ..............IEC 60529: IP20 Electromagnetic Compatibility (EMC) International ............IEC 61326-1: Basic Electromagnetic Environment CISPR 11: Group 1, Class A Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio frequency energy that is necessary for the internal function of the equipment itself.
  • Page 26: Detailed Specifications

    ESA614 Getting Started Manual Detailed Specifications Test Standard Selections ........ANSI/AAMI ES-1 Voltage Ranges (Mains voltage) ........90.0 V to 132.0 V ac rms 180.0 V to 264.0 V ac rms Range (Point-to-point voltage) 5000 m ............0.0 V to ≤150 V ac rms 2000 m ............
  • Page 27 Electrical Safety Analyzer Detailed Specifications Patient Load ............AAMI ES1-1993 Fig. 1 Crest factor ............≤3 Ranges ..............0.0 μA to 199.9 μA 200 μA to 1999 μA 2.00 mA to 10.00 mA Accuracy DC to 1 kHz ............. ±(1 % of reading + (1 μA or 1 LSD, whichever is greater)) 1 kHz to 100 kHz ..........
  • Page 28 ESA614 Getting Started Manual ECG Performance Waveforms Accuracy Frequency ............±2 % Amplitude ............±5 % of 2 Hz square wave only, fixed @ 1 mV Lead II configuration Waveforms ECG Complex ..........30, 60, 120, 180, and 240 BPM Ventricular Fibrillation Square wave (50 % duty cycle) ......