Accessing the Analyzer Tests

The Analyzer uses a series of menus to access various
Analyzer functions and setup variables. As shown in
Figure 4, the Analyzer indicates three different defibrillator
tests (Energy, Sync, and Charge Time) along the bottom
of the display. An Exit selection is also indicated as a way
of backing out of the defibrillator tests. Pressing a softkey
(F1 through F5) under a specific test will cause that test to
be selected.
Figure 4. Defibrillator Menu
Some menu selections reveal a list of selectable items by
displaying K to the right of the presently selected item.
See Figure 5. To change the selection, press either G or
H to scroll through the possible selections. Once the
desired selection appears, press the function softkey and
K disappears from the display.
8
Figure 5. Cursor Navigation Example

Analyzing Defibrillators

There are three main defibrillator test functions to evaluate
a defibrillator's performance: Energy, Synchronization, and
Charge Time. To set the Analyzer for defibrillator testing,
press M.
The Analyzer's defibrillator input connectors are designed
to be used with test leads or adapter plates when testing
fak02.eps
defibrillators using external paddles.
Connect the defibrillator to the Analyzer as shown in
Figure 6.

Testing Energy Levels

Press the softkey labeled Energy to enter the energy test
procedure. As shown in Figure 7, the Analyzer has a
waveform selection already set. Either the waveform
characteristic is off or it was the last one setup from a
previous defibrillator test.
fak03.eps