General Specifications

Battery Power ........................................................ 4 x Panasonic HHR210AB NiMh 2000 mAh batteries
Charger
Operating Voltage Range ................................... 100 V ac to 240 V ac
Supply Frequency ............................................... 50 Hz / 60 Hz
Supply Power ...................................................... <20 VA
Size (HxWxD) ......................................................... 30 cm x 17 cm x 10 cm (12 in x 8 in x 4 in)
Weight .................................................................... ~1.2 kg (2.7 lb)
Temperature
Operating ............................................................ 15 °C to 30 °C (59 °F to 86 °F)
Storage ............................................................... -20 °C to +40 °C (-4 °F to +104 °F) when drained of all liquid.
Humidity ................................................................. 10 % to 90 % non-condensing
Altitude ................................................................... 0 meters to 2000 meters (6500 feet)
Safety ..................................................................... IEC 61010-1: Overvoltage category II, Pollution Degree 2
Electromagnetic Environment ............................. IEC 61326-1: Basic
Emissions Classification...................................... IEC CISPR 11: Group 1, Class A. (Group 1 have intentionally
FCC ......................................................................... CFR47: Class A Part 15 subpart B
Storage of Results ................................................ Test results stored for later viewing, printing or transfer to PC. Typical
Power Down .......................................................... The results of tests in progress will be saved in the case of accidental
Computer Control ................................................. The Product can be fully controlled from a PC using HydroGraph

Performance Specifications

Average Flow Rate Measurement
Technique ........................................................... Flow is calculated by measuring volume over time
Range ................................................................. 0.5 ml/h to 1000 ml/h
Accuracy ............................................................. 1 % of reading ±1 LSD for flows of 16 ml/h to 200 ml/h for volumes
Max test duration ................................................ 10 hours on battery
Volume Measurement
Technique ........................................................... Volume is measured directly by the measuring module in minimum
Range ................................................................. 0.06 ml to 999 ml
Accuracy ............................................................. 1 % of reading ±1 LSD for flow rates of 16 ml/h to 200 ml/h for volumes
Max test duration ................................................ 10 hours on battery
Pressure Measurement
Technique (Occlusion test) ................................. Direct measurement of pressure at the inlet port
Range ................................................................. 0 psi to 45 psi and equivalent in mmHg, Bar and kPa
Accuracy ............................................................. 1 % of Full Scale ±1 LSD under laboratory conditions
Max test duration ................................................ 30 minutes
generated and/or use conductively coupled radio-frequency energy
which is necessary for the internal functioning of the equipment itself.
Class A equipment is suitable for use in non-domestic locations and/or
directly connected to a low-voltage power supply network.)
practical capacity: 100 tests
power down
software
over 20 ml; otherwise 2 % of reading ±1 LSD for volumes over 10 ml
under laboratory conditions
sample sizes of 60 µl
over 20 ml. Otherwise 2 % of reading ±1 LSD for volumes over 10 ml
under laboratory conditions