Synchronization Time Test

The Synchronization Time test determines the ability of
the defibrillator to synchronize the discharge of its output
pulse with the simulated ECG waveform generated by the
Analyzer. Synchronization Time test measurements are
Measurement
Sync Time
Energy
Peak Voltage
Peak Current
Table 8. Charge Time Test Custom Parameters
The energy (in Joules) used for the test
The field user is prompted to set the defibrillator to this value.
The ECG waveform that the Analyzer should simulate during the test
Table 9. Synchronization Time Test Measurements
Unit of Measure
seconds
The delay between the top of the ECG wave and the discharging of the
defibrillator pulse
Joules
Amount of energy discharged by the defibrillator
Volts
Peak voltage detected during the discharge
Amperes
Peak current detected during the discharge
Defibrillator/Transcutaneous Pacemaker Analyzer
Using Defibrillator Test Elements
Description
listed in Table 9. The Synchronization Time test custom
parameters are shown in Table 10.
This test element supports the Auto Advance
Test Guide Setting.