ECG Pacer Interactive Test

The ECG Pacer Interactive test simulates a patient
response to a pacemaker. Table 17 lists the ECG Pacer
Interactive custom test parameters.
68
Table 16. Pacer Sensitivity Test Custom Parameters
Specifies where the pacer leads are attached on the Impulse 7000DP
Choose between pacer jacks or defibrillator jacks.
Defines load used for the test
If defibrillator jacks are used for the test, a fixed 50 Ω load is used.
The brand of defibrillator/pacer being tested can be specified to optimize the accuracy of the test.
Expected pacer rate; field user prompted to set pacemaker to this rate
Expected pacer amplitude; field user prompted to set the pacemaker output current to this value
The type of ECG waveform output to generate during the test
The width of each waveform pulse
The ECG waveform polarity that can be specified
Description
Note
This is a visual / audible test; the Analyzer takes
no measurements during this test.