ECG Noise Immunity Test
The ECG Noise Immunity test checks the ability of the
ECG monitor to reject AC line frequency noise. The
custom parameters for this test are listed in table 23.
Parameter
Noise Amplitude
Line Frequency
ECG Wave
Using Battery Performance Test
Elements
Battery Capacity Test
The Battery Capacity test is used to check whether a
battery-powered defibrillator can deliver a certain number
Table 23. ECG Noise Immunity Test Custom Parameters
Defibrillator/Transcutaneous Pacemaker Analyzer
Using Battery Performance Test Elements
This is a visual / audible test; the Analyzer takes
no measurements during this test.
Description
Indicates the amplitude for the waveform
Selects 50 Hz or 60 Hz line noise simulation
Option to include an ECG Wave during the test
of discharges per minute and whether the charge time
remains adequate throughout the test. Tables 24 and 25
list the Battery Capacity test measurements and custom
parameters.
Note
75