Fluke Biomedical Impulse 6000D User Manual page 53

Defibrillator analyzer, defibrillator/transcutaneous pacer analyzer
Hide thumbs Also See for Impulse 6000D:
Table of Contents

Advertisement

Pulse Rate
Range..................................................................5.0 to 800 PPM
Accuracy..............................................................±(0.5% of reading + 0.1 PPM)
Pulse Width
Range..................................................................1.00 to 100.0 ms
Accuracy..............................................................±(0.5% of reading + 0.01 ms)
Energy
Range..................................................................1 µJ to 2.00 J
Accuracy..............................................................±(4% of reading + 10 µJ)
Demand and Asynchronous Mode Test
Input Pacer pulse rates.........................................30 to 200 PPM
ECG NSR wave
Rate.....................................................................10 to 300 (by 1) BPM
Amplitude ............................................................1 mV
Underdrive rate....................................................10 BPM minimum
Overdrive rate......................................................300 BPM maximum
Defibrillator/Transcutaneous Pacemaker Analyzer
Specifications
39