Impulse 6000D, 7000DP
Users Manual
Arrhythmia Wave Test
This test is typically used to verify the shock advisory
capability of a defibrillator in response to various
arrhythmia waveforms. Table 20 lists test custom
parameters.
Parameter
Arrhythmia
Arrhythmia Type
Poly VTach Type
Mono VTach Rate
ECG Amplitude
Duration
Atrial Width, Amplitude,
and Polarity
Ventricular Width,
Amplitude, and Polarity
72
Table 20. Arrhythmia Wave Advisory Test Custom Parameters
Selects the category of arrhythmia waveform to use for the test
Selects the type of arrhythmia to simulate
Provides a choice of 5 types of Poly VTach waveforms when this type of arrhythmia is selected
BPM value is required when Mono VTach Rate is selected.
The voltage amplitude to use for the arrhythmia simulation
How long the simulation should last. Setting can range from 1 to 60 seconds. With checkbox
Indefinite selected, duration is ignored, and the test must be manually stopped by the field user
(Stop in TEST GUIDE toolbar).
These atrial parameters are needed for some TV Paced arrhythmia simulations.
These ventricular parameters are needed for some TV Paced arrhythmia simulations.
Note
This is a visual / audible test; the Analyzer takes
no measurements during this test.
Description