Defib Pulse Repetition Test

The Defib Pulse Repetition test is used to determine if a
battery-powered defibrillator can deliver a specified
Measurement
Unit of Measure
Energy
Joules
Parameter
Preset Energy
Number of Pulses
Number of Minutes
Table 26. Defib Pulse Repetition Test Measurements
Amount of energy discharged by the defibrillator
The test stores multiple energy measurements – one per discharge.
Table 27. Defib Pulse Repetition Test Custom Parameters
The energy (in Joules) used for the test
The field user is prompted to set the defibrillator to this value.
The number of discharges to be completed during the test
The maximum time allowed for this test to be completed
Defibrillator/Transcutaneous Pacemaker Analyzer
Using Battery Performance Test Elements
number of discharges within a specified amount of time.
Tables 26 and 27 list the Defib Pulse Repetition test
measurements and custom parameters.
Description
Description
77

Hide quick links: