Defibrillator Analyzer Specifications

Energy Output Measurement
Compatible Defibrillator Waveshapes.................Lown, Edmark, Trapezoidal, DC Bi-phasic, and AC Pulsed Bi-phasic
AC Pulsed Bi-Phasic waveform has not been approved in the United States.
Autoranged Measurement ....................................0.1 to 600 J
Accuracy
0.1 to 360 J .....................................................±(1 % of reading + 0.1 J)
360 to 600 J ....................................................±(1 % of reading + 0.1 J), typical
For Pulsed Bi-Phasic defibrillator, specified accuracy is
Load resistance
Resistance...........................................................50 Ω
Accuracy..........................................................±1 %, non-inductive (<2 μH)
Pulse trigger level .................................................20 V
Pulse width
Range..................................................................1.0 to 50.0 ms
Accuracy..............................................................±0.1 ms
Voltage
Range..................................................................20 to 5000 V
Accuracy..............................................................±(1 % of reading + 2 V)
Current
Range..................................................................0.4 to 100.0 A
Accuracy..............................................................±(1 % of reading + 0.1 A)
Defibrillator/Transcutaneous Pacemaker Analyzer
Note
Note
±
(1.5 % of reading + 0.3 J) on both ranges.
Specifications