Fluke Biomedical Impulse 6000D User Manual page 75

Defibrillator analyzer, defibrillator/transcutaneous pacer analyzer
Hide thumbs Also See for Impulse 6000D:
Table of Contents

Advertisement

Measurement
Energy
Peak Voltage
Peak Current
Pulse Width 50%
Pulse Width 10%
Parameter
Preset Energy
ECG Waveform
Table 5. Energy Measurement Test Measurements
Unit of Measure
Joules
Amount of energy discharged by the defibrillator
Volts
Peak voltage detected during the discharge
Amperes
Peak current detected during the discharge
milliseconds
The width of the pulse at 50% of its peak
milliseconds
The width of the pulse at 10% of its peak
Table 6. Energy Measurement Test Custom Parameters
The energy (in Joules) used for the test
The field user is prompted to set the defibrillator to this value.
The ECG waveform that the Analyzer should simulate during the test
Defibrillator/Transcutaneous Pacemaker Analyzer
Using Defibrillator Test Elements
Description
Description
61