the BPM is outside the limits specified in the test
procedure, the test is automatically marked as failed.

Arrhythmia Wave Test

The Arrhythmia Wave test typically verifies the shock
advisory capability of a defibrillator in response to various
arrhythmia waveforms.
To run an Arrhythmia Wave Test:
1.
Connect the device under test (DUT) to the Analyzer.
Set up the shock advisory on the defibrillator, if
applicable.
2.
Click Start in the TEST GUIDE toolbar. Wait for
activity to appear on the ECG monitor and verify the
shock advisory was correct. The test concludes
automatically after a few seconds, as specified by the
test procedure.
3.
If the test is set to run indefinitely, click Stop in the
TEST GUIDE toolbar at any time to end the test.
4.
Click the checkbox Test passed or the checkbox Test
failed to record the test result based on what was
observed on the monitor.
Defibrillator/Transcutaneous Pacemaker Analyzer
ECG Waveform Simulation Tests

Performance Wave Simulation

The Performance Wave simulation tests the integrity of a
defibrillator monitor using a variety of additional waveform
shapes such as square, triangle, sine, and pulse.
Refer to the Normal Sinus Wave Simulation test
procedure for directions in running this test.

ECG R-Wave Test

The ECG R-Wave (Peak Detection) test determines if the
defibrillator can detect an R Wave at a given threshold of
width and amplitude. Beats per minute can range
between 30 and 250 BPM. This test is set up to test a
single R-wave width and amplitude, or it can cycle
through several widths or several amplitudes.
To run an ECG R-Wave test:
1.
Follow the directions provided by the DUT equipment
manufacturer to connect the ECG leads from the
ECG monitor to the Analyzer.
2.
Click Start in the TEST GUIDE toolbar.
3.
Check whether the ECG monitor responds correctly
based on the settings being used for the test.
4.
If the test duration is set to run indefinitely, click Stop
in the TEST GUIDE toolbar to conclude the test.