Specifications - Fluke Biomedical Victoreen 07-494 Operator's Manual

Wide-range digital kvp meter
Table of Contents

Advertisement

Victoreen 07-494
Operators Manual

1.2 Specifications

Range
Resolution
Accuracy
mAs Requirements
Reproducibility
Filtration Effects
Calibration Frequency
Battery
Display
LED Indicators
Controls
Connectors
Dimensions
(H x W x D)
Weight
Environmental Specifications
1-2
Low: 50 kVp to 90 kVp
High: 80 kVp to 150 kVp
0.1 kVp
± 3% or 3 kV, whichever is greater (tungsten target x-ray tube
with 4.5 mm Al total filtration)
See Figure 1-2
± 1 kV or 1%, whichever is greater
50 kVp - 90 kVp range: less than 3.5% with 5.3 mm aluminum
added
80 kVp – 150 kVp range: less than 3% with 8 mm aluminum
added
One (1) year
Type: 9 V alkaline, NEDA 1604 (IEC 6F22) or alkaline equivalent
Life: 75 hr. typical
3 ½ digit, 0.5" high LCD with LO BAT indicator;
Below Range kVp: Flashing middle decimal point
Above Range kVp: Flashing left decimal point
Above Range mAs: Flashing right decimal point
Below Range mAs: Flashing right decimal point and 00.0
display
Auto Reset: New exposure detected, last kVp reading cleared
Range Fault: Range select sliding filter not properly engaged
Power Switch: ON/OFF
Phase Selection Switch: Single phase/3 phase
Filter Range Selection Switch: 50 to 90 kVp or 80 to 150 kVp
BNC connector for waveform output
2.5 in. x 8 in. x 6 in.
(6.4 cm x 20.3 cm x 15.2 cm)
2.4 lbs. (1.1 kg)
Temperature:
Operating: 59 to 104° F (15 to 40° C)
Storage: - 0 to 122° F (-18 to +55° C)
Relative Humidity:
Operating: 0 to 90% non-condensing