Appendix B - Impulse 6000D/7000Dp Test Templates; Introduction; Creating Test Templates - Fluke Biomedical Impulse 6000D User Manual

Defibrillator analyzer, defibrillator/transcutaneous pacer analyzer
Hide thumbs Also See for Impulse 6000D:
Table of Contents

Advertisement

Introduction

This chapter introduces the template capabilities of the
Impulse 6000D/7000DP Plug-In and provides guidance
for customizing test templates.

Creating Test Templates

Create, modify, and review test templates using the Ansur
Main Application window as a template editor. The
Impulse 6000D/7000DP Plug-In provides 16 test
elements that are used to build new test procedures.
These are accessible in Test Explorer and are coded as
follows:
Light blue icon – the Analyzer automatically provides
test result data to Ansur as the test is completed.
Impulse 6000D/7000DP Test Templates
To build a test template, take the following actions,
beginning from the Main Application window:
1.
Yellow icon – resultant data must be manually
entered into Ansur by the user.
Drag a test element from the Test Explorer (left
pane) into the Test Template (right pane), as
displayed in Figure 2-32. Clicking the test element in
the Test Template highlights the test element and its
properties. In this illustration, the highlighted element
is the Impulse 6000D/7000DP Energy
Measurement Test, the first test step to be
performed.
Appendix B
53