Defibrillator Tests; Energy Measurement Test - Fluke Biomedical Impulse 6000D User Manual

Defibrillator analyzer, defibrillator/transcutaneous pacer analyzer
Hide thumbs Also See for Impulse 6000D:
Table of Contents

Advertisement

Defibrillator Tests

The Impulse 6000D/7000DP Plug-in allows testomg of
defibrillator performance using a PC running the Ansur
software. At the conclusion of each test procedure, Ansur
collects the results of the tests to display or to store on a
PC.

Energy Measurement Test

The energy tests verify the accuracy of the energy
delivered by the defibrillator.
To run an Energy Measurement test:
1.
First connect the defibrillator to the Analyzer (refer to
the Impulse 6000D/7000DP Users Manual for
connection instructions) and set the defibrillator to
the energy setting displayed in the information block
in the right pane of the TEST GUIDE window. Figure
1-31 shows an energy setting with both high and low
limits (in Joules).
2.
Click Start in the TEST GUIDE toolbar.
3.
Charge the defibrillator.
4.
Ansur software configures the Analyzer for the
defibrillator test, indicating the status in the Test
Defibrillator/Transcutaneous Pacemaker Analyzer
results pane. Wait for Ansur to finish configuring the
Analyzer.
5.
When configuration is complete, a window displays
the prompt to discharge the defibrillator.
6.
Discharge the defibrillator. Test results appear in the
Test results pane immediately.
To view a graph of the discharge curve, as shown in
Figure 31, click the graph tab located along the left side of
the TEST GUIDE.
Defibrillator Tests
43