Parameter
Response Waveform
Rate
Duration
Response Threshold
Table 17. ECG Pacer Interactive Test Custom Parameters
The type of patient response to simulate:
Demand: Normal sinus rhythm at specified rate. Pacer in demand mode can interact so it
paces when the normal rate is too slow. Heart responds to pacer pulses at or above
threshold.
Asystole: No heartbeat, but heart responds to pacer pulses at or above threshold.
Non-capture: Same as Asystole, but heart fails to respond to one out of every ten pacer
pulses.
Non-function: No heartbeat and no response to pacing.
Beats per minute to use for the Demand Response Waveform
How long the simulation should last
Setting can range from 1 to 60 seconds. If the "Indefinite" checkbox is checked, the duration is
ignored, and the test must be manually stopped by the field user (Stop in TEST GUIDE toolbar).
The level of pacing current required to generate a heart response
Defibrillator/Transcutaneous Pacemaker Analyzer
Using Pacemaker Test Elements (Impulse 7000DP only)
Description
69