Weekly Maintenance - Siare Siaretron 4000 Service Manual

Icu ventilator, turbine-driven ventilation
Hide thumbs Also See for Siaretron 4000:
Table of Contents

Advertisement

3.4

Weekly maintenance

Gas supply " OFF "
Power supply " OFF "
Turbine air filters
Open the right front door of the trolley, gas side.
Remove the cover (and filters, inside the cover).
Replace O-Ring, code: M10003193 from turbine
plastic interface ( please see note, CAUTION ).
Remove the turbine air filters.
Replace with a new turbine air filters ( code,
M53553009 ).
Reassemble the cover and filters on
Siaretron 4000.
CAUTION
Maintenance operations that should be carried out several times a month or
according to the local procedures and reference standards.
For more details see on chapter 7 Maintenance.
CAUTION
The O-Ring, code: M10003193, have to replaced 1 time to six month.

Hide quick links:

Advertisement

Table of Contents
loading