Service Procedures; Opening The Ventilator Crankcase - Siare Siaretron 4000 Service Manual

Icu ventilator, turbine-driven ventilation
Hide thumbs Also See for Siaretron 4000:
Table of Contents

Advertisement

3.8

Service procedures

3.8.1 Opening the ventilator crankcase

Gas supply " OFF "
Power supply " OFF "
Unscrew the 8 fastening screws of the
rear ventilator crankcase
Remove the crankcase from its seat
Unscrew the 5 fastening screws of the
rear ventilator work surface
Pull the work surface horizontally to
remove it from its seat
WARNING! Patient injury hazard
The procedures described are critical operations and must be carried out only by
authorised staff as they might affect the equipment's safety and proper operation.
WARNING! Clinician injury hazard
Please turn off the ventilator and disconnect it from the mains and from the medical
gas distribution line before carrying out any service procedure.
WARNING! Clinician injury hazard
Engage the trolley's brakes, to make sure that it won't move during service phases.