5.3.28 PSV (Pressure Support Ventilation)
Press the START key and make sure that the device operates as intended
(according to the relative operating mode). Monitored parameters check.
Based on the time set in parameter MENU – SETUP – APNOEA TIME (sec),
Apnoea Back
Up
check the proper mode activation.
The system will generate an acoustic (visual) signal notifying the clinician
on the alarm condition.
Press the silence alarm soft key and make sure that the system returns to
the relative operating mode.
BK
The device provides ventilation, based on the set bk parameters.
parameters
Ventilator monitored parameters check.
check
5.3.29 PSV-TV (Volume Targeted Pressure Support Ventilation) operating mode
Press the START key and make sure that the device operates as intended
(according to the relative operating mode). Monitored parameters check.
Based on the time set in parameter MENU – SETUP – APNOEA TIME (sec),
Apnoea Back
Up
check the proper mode activation.
The system will generate an acoustic (visual) signal notifying the clinician
on the alarm condition.
Press the silence alarm soft key and make sure that the system returns to
the relative operating mode.
BK
The device provides ventilation, based on the set bk parameters.
parameters
Ventilator monitored parameters check.
check
The system will activate the "Low Pressure" alarm after 20
sec. from the moment when the patient stops breathing.
The system will activate the "Low Pressure" alarm after 20
sec. from the moment when the patient stops breathing.