9.1
Equipment
This paragraph illustrates what specific equipment is necessary to correctly and safely
carry out the maintenance and service procedures planned for Siaretron 4000 lung
ventilator.
9.1.1 Measuring instruments
Electrical safety analyser
Digital multimeter
Electronic test for lung ventilators to measure:
o
Pressure, Flow, Volume, Respiratory rate, Concentration of O 2
Mechanical pressure gauge of 0-80 cm H 2 0 and Mechanical pressure gauge of 0-10
Bar
Lung simulator with variable compliance and resistance
9.1.2 Equipment
Medical air compressor
Series of hex keys
Series of tube wrenches
Series of 1.5 to 10 mm Allen wrenches
Slotted and cross-head screwdrivers
Screwdriver with interchangeable bits:
o
2.5 mm shear head hexagonal bit
o
3 mm shear head hexagonal bit
Pliers, slip joint pliers, nose pliers
Tweezers
1 mm seeger pliers for indoors and outdoors
Rubber hammer
Cutting plier
Electrician's scissors, wire stripper
Cutter
Vaseline grease
Thread seals (Teflon liquid) (LOXEAL 58-11)
Retaining adhesive (LOXEAL 83-21)