Dräger Infinity Series User Manual page 4

Configured monitoring
Hide thumbs Also See for Infinity Series:
Table of Contents

Advertisement

Infinity Gamma Series Software Release Notes
Wireless network operation requires special configurations of the monitor-
!
ing network and the M
you experience problems with wireless network operations, contact your
Service personnel.
When moving and assigning a wireless monitor to a different central sta-
!
tion, the original central station may emit a brief network error tone and
display an Offline message instead of the message Bed Disconnected.
However, there is no disruption of network monitoring and the Offline
message clears as soon as you assign a new bed to the central viewport.
When you change the units of measure at the bedside and the central sta-
!
tion is showing the monitor's bed view, you must first exit the central bed
view, before the change of units appears at the central station.
For network and card data transfer:
!
— Occasionally, after a data transfer from an Infinity Delta Series monitor,
you may see three or four ST trends instead of the two ST leads monitored
by the Infinity Gamma Series monitor.
!
For network data transfer only:
— If ST is enabled, the ST data transferred from an Infinity Delta Series
monitor (VE0) or from a M
I and II, regardless of the ST leads selected on the Infinity Gamma Series
monitor. Note: Other ST data will be permanently lost.
— After a network transfer of telemetry data to an Infinity Gamma Series
monitor, ST trend points may appear two minutes apart.
— The IBP data transferred from an Infinity Delta Series monitor (VE0) to
an Infinity Gamma Series monitor is labelled GP1 and GP2.
NBP parameter values transferred from an Infinity Gamma Series monitor
!
to an Infinity Delta Series monitor will be displayed in the trend graphs
rather than in the trend tables of the destination monitor.
On rare occasions, a docked monitor may reset when entering the
!
Transfer Menu under certain network conditions. The monitor returns
to the state prior to the reset within 30 seconds.
When power-cycling the monitor or admitting a new patient, saved moni-
!
toring settings may occasionally return to default settings. Check monitor-
ing settings after these events.
!
If the Scio module is unable to measure the concentration of N
itor may enter the error code *A* (artifact) instead of *F* (failure) into the
trend storage.
Page iv
Software Version VF4
V
W
ULTI
IEW
ORK
V
ULTI
IEW
Infinity Gamma Series
S
(a service function). If
TATION
Telemetry System (VE0) is ST lead
O, the mon-
2
VF4

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Infinity gamma seriesGamma xl

Table of Contents