Alarm Validation - Dräger Infinity Series User Manual

Configured monitoring
Hide thumbs Also See for Infinity Series:
Table of Contents

Advertisement

Alarm Validation

The alarm validation feature minimizes nuisance alarms for tran-
sient conditions. When a parameter value falls below or rises
above its current alarm limits, the monitor waits a predetermined
time before triggering the alarm. If the parameter value returns to
within the upper or lower alarm limit before the end of the alarm
validation period, the monitor does not trigger the alarm.
The alarm validation period is part of the software (see table
below) and cannot be modified. The monitor triggers the alarm at
the end of the validation period. Parameters have the following
alarm validation periods:
Parameter-Specific Alarm Validation Periods
Parameter
ECG/Heart rate (HR)
Pulse rate (PLS)
ST Segment Analysis*
Respiration rate (Rsp)
Arterial Oxygen Saturation (SpO
etCO
*, iCO
*, T, PVC, Arrhythmia
2
2
Events
Scio Multigas*
RRc*
Invasive blood pressure (ART, PA,
CVP, ICP, GP1, GP2*)
Non-invasive blood pressure (NBP)
*Available only if option is enabled
Page 5-10
Upper Limit
(in seconds)
60
)
2
Infinity Gamma Series
Lower Limit
(in seconds)
2
4
8
4
0
0
8
4
0
0
10
60
10
10
0
0
10
4
0
VF4

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Infinity gamma seriesGamma xl

Table of Contents