Multigas
(continued)
Message: iO
> UL
2
Message: iO
< LL
2
Parameter value: <value>
Message: etO
> UL
2
Message: etO
< LL
2
Parameter value: <value>
Message: i <agent> > UL
Message: i <agent> < LL
Parameter Value: <value>
Message: et <agent> > UL
Message: et <agent> < LL
Parameter value: <value>
Message: <agent> detected
Parameter value: <blank>
Message: Mixed Agents
Parameter value: ***
Message: <parameter>
reduced accuracy
Parameter value: <value>
Message: Multigas
Unplugged
Parameter value: ***
VF4
Alarms and Messages
The inspired/expired
concentrations fall
outside the current
upper or lower alarm
limits.
The user has specified
an agent (Agent Over-
ride), but Scio is
detecting a different
agent.
Scio is sensing more
than one agent in the
breathing circuit.
Scio reports reduced
or unknown accuracy
even after completion
of warm-up period.
No connection or
faulty connection
between monitor and
Scio module.
Infinity Gamma Series
• Observe the patient;
treat, if necessary.
• Adjust alarm limits.
• Turn Agent Override
off.
• After a change of
agent, wait until all
traces of the first agent
have been cleared out
of the breathing circuit.
• Observe the patient;
treat, if necessary.
• Check connections.
• Disconnect and recon-
nect the Scio module.
• Power-cycle the moni-
tor or undock/redock
the monitor from the
Docking Station.
• Contact DrägerService.
• Check connections and
cables.
• If message persists,
contact DrägerService.