Dräger Infinity Series User Manual page 134

Configured monitoring
Hide thumbs Also See for Infinity Series:
Table of Contents

Advertisement

Multigas
(continued)
Message: Multigas H/W fail-
ure
Parameter value: ***
Message: Multigas initializa-
tion
Parameter value: <blank>
Message: Multigas warming
up
Parameter value: <values>
Message: Multigas zero in 1
minute
Parameter value: <values>
Message: Multigas zero in
progress
Parameter value: <values>
Message: Multigas zero OK
Parameter value: <values>
Message: Multigas zero
failed
Parameter value: ***
Message: Multigas Tubing
Blocked
Parameter value: ***
Page 5-30
Loss of communica-
tion or hardware prob-
lem detected.
Scio start-up mode.
Scio warm-up period:
Reduced accuracy.
Scio will start the auto-
matic zeroing process
in one minute.
Scio is undergoing
automatic zeroing pro-
cess. During zeroing
(< 25s), the monitor
does not update the
displayed gas values.
Scio zeroing was suc-
cessful.
Scio zeroing was not
successful.
Scio reports an occlu-
sion or pneumatic
problem.
Infinity Gamma Series
• Check connections.
• Disconnect and recon-
nect the Scio module.
• Power-cycle the moni-
tor or undock/redock
the monitor from the
Docking Station.
• Contact DrägerService.
• Informational message.
• Displayed values might
not be accurate. Scio
reaches full accuracy
after a warm-up period
of about 5 minutes.
• To delay zeroing for 5
minutes, click on
Autozero Delay in the
Multigas Setup menu
(see the chapter Multi-
gas).
• Wait until zeroing is
completed. During
zeroing, displayed gas
values might not be
accurate.
• Informational message.
• Verify that the environ-
ment does not contain
contaminants.
• Check for leaks or
occlusions.
• Contact DrägerService.
• Check all catheters for
occlusion.
• If necessary, replace
catheters.
VF4