etCO2 Alarms

The monitor alarms for the following etCO
etCO
!
breath).
iCO
!
breath, or rebreathing).
RRc alarm (significant changes in the respiratory rate,
!
derived from the capnogram waveform).
RRc Apnea alarm (apnea conditions based on the respiratory
!
rate derived from the capnogram waveform).
In addition, the monitor alarms for unplugged cables, weak sig-
nals, and in the absence of periodic measurement updates.
You can set etCO
modes within the following range:
5 to 95 mmHg for the etCO
!
4 to 10 mmHg for the iCO
!
5 to 145 breaths per minute for the RRc alarm (mainstream).
!
5 to 65 breaths per minute for the RRc alarm (sidestream).
!
Turn etCO
Limits table (see the chapter Alarms and Messages).
VF4
alarm (critical levels of CO
2
alarm (critical levels of CO
2
upper and lower limits for all monitoring
2
alarms on or off and set the alarm limits in the Alarm
2
Infinity Gamma Series
End-Tidal CO2
parameters:
2
saturation in the exhaled
2
saturation in the inhaled
2
alarm (0.7 to 12.7 kPa or %).
2
alarm (0.5 to 1.3 kPa or %).
2
Page 13-19