Dräger Infinity Series User Manual page 55

Configured monitoring
Hide thumbs Also See for Infinity Series:
Table of Contents

Advertisement

4. With the battery pushed into the left side of the battery com-
partment, press the right side of the battery into the clip at the
right side.
5. Insert the left side of the battery door into the hinges, and
swivel the door closed until the locking tab snaps into place.
6. Press the
light indicator in the key should light up.
7. Wait until the end of the self-tests and verify that the battery
gauge appears at the bottom of the display.
NOTE: The battery gauge appears only when you operate the
monitor on battery power.
To prolong battery life, we recommend the following:
Use the AC adapter whenever possible.
!
Keep the battery charged.
!
STEPS: Charging the Battery with the Monitor
The battery that is delivered with a new monitor needs to be fully
charged before monitoring. When connected to line power via the
AC Adapter, IDS, or CPS, the monitor automatically charges the
battery and the green Battery Charger LED lights up (the monitor
can be turned on or off).
To charge the battery:
1. Connect the monitor to line power and let the lead acid bat-
tery charge for 5½ hours, the lithium ion battery for 8 hours.
(The Battery Charging message appears intermittently.)
2. Wait until the battery is fully charged before monitoring a
patient. (The Battery Charging message no longer appears.)
NOTE: If the battery does not charge properly, the message Bat-
tery Charger Error appears. Contact your Biomedical technician to
replace the battery.
VF4
fixed key to turn the monitor on. The green
Power
Infinity Gamma Series
Monitor Setup
Page 2-7

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Infinity gamma seriesGamma xl

Table of Contents