Dräger Infinity Series User Manual page 132

Configured monitoring
Hide thumbs Also See for Infinity Series:
Table of Contents

Advertisement

End Tidal CO
(etCO
2
Message: etCO
Tubing
2
Blocked
Parameter value: ***
Message: etCO
Tubing
2
Leak
Parameter value: ***
Message: etCO
2
Hardware Failure
Parameter value: ***
Multigas
Displayed Message
Message: iCO
/etCO
2
Range
Parameter value: +++
Message: iO
/etO
2
2
Range
Parameter value: +++
Message: i/et <agent> Out of
Range
Parameter value: +++
Page 5-28
), Inspired CO
2
2
Sidestream tubing
obstructed, or filter is
clogged.
Sidestream tubing has a
leak.
etCO
communication or
2
hardware failure.
Possible Cause
Out of
The inspired/expired
2
concentrations fall
outside the monitor's
display range.
Out of
Infinity Gamma Series
(iCO
), and RRc
(continued)
2
• Clear the blockage in
the tubing.
• Replace the pod.
• Change the tubing.
• Check etCO
nection.
• Check all tubing
• Disconnect the etCO
pod, then reconnect it.
If the message persists,
contact your Biomed.
Suggested Action
• Observe the patient;
treat, if necessary.
• Check connections.
• Disconnect and recon-
nect the Scio module.
• Power-cycle the moni-
tor or undock/redock
the monitor from the
Docking Station.
• Contact DrägerService.
pod con-
2
2
VF4