Adjustment Procedures

Adjustment Interval

Adjustment After Repair

Adjustment

MSO70000/C, DSA70000B/C, DPO70000B/C, DPO7000, MSO5000, DPO5000 Series
This chapter contains information about instrument adjustment.
The voltage and timing references inside the instrument are very stable over time
and should not need routine adjustment.
If the instrument fails the performance tests in the Specifications and Performance
Verification manual, adjustment may be required. This manual was provided with
your product, and can also be found at www.tektronix.com/manuals by searching
for your product.
If periodic calibration is one of your requirements, a general rule is to verify
performance and make adjustments (only if needed) every 2000 hours of operation
or once a year if the instrument is used infrequently.
After removal and replacement of a module, you must perform the Performance
Verification procedure, found in the Specifications and Performance Verification
manual. This manual was provided with your product, and can also be found at
www.tektronix.com/manuals by searching for your product.
If the instrument fails the Performance Verification tests, it must be returned to
Tektronix for calibration.
Adjustment can be performed only by a Tektronix Service Center. See Contacting
Tektronix, following the title page in this manual, for information on contacting
Tektronix Service Support.