3.2 System Installation Procedure

No.
Step
Unpack and check the
1
product's constituent parts
Affix the insulation sheets to
2
the sensor unit
Install the imaging unit
3
Connect the imaging unit and
power box
4
Connect the power box and the
5
control PC
Attach and connect the
6
operation unit
Adjust the alignment
7
Install the phototimer
8
Start the system
9
Identifying the imaging units
and setting the number of units
10
to be connected (inputting the
sensor serial numbers)
Install control PC unit serial
number.
11
Check the software program's
version
12
Check date and time
13
Connect the power box and
14
X-ray generators
Adjusting the timing with the
15
X-ray generators
Perform calibration imaging
16
2. Installation
Conditions and checkpoints
- There are no missing parts, damage,
scratches, or other imperfections.
- Affix to the specified locations on the
imaging unit without making any raised or
bent sections.
- Install and secure in place in the Bucky unit,
stand or table. Obtain the imaging unit
fixture pieces from a distributor.
- Be careful not to drop the imaging unit.
- The cables must be routed so that no
unreasonable loads are placed on them.
- R at stationary parts (radius of cable inner
circumference) = 25 mm or more
- R at moving parts (radius of cable inner
circumference) = 50 mm or more
- The cables must be routed in such a way that
no unreasonable loads are brought to bear
upon them.
- The cables must be routed in such a way that
no unreasonable loads are brought to bear
upon them.
- Be careful not to scratch the imaging unit.
- Install in the Bucky unit, stand or table.
- Be careful not to scratch the sensor.
- The compatibility of the sensor unit and the
control PC must be checked on the
compatibility list, and the software program
must be installed or upgraded as required.
- Date and time must be changed according to
the area where the instrument is installed.
- The manufacturer of the X-ray generators
must be asked to handle the connections
with the generators.
- This is generally not necessary.
- No errors must be displayed.
- Calibration imaging must be performed by
service staff only.
- 9 -
Reference
"Identifying the
Sensor Unit" in
"Setting"
"Entering
Control PC
Serial Number"
in "Setting"
"Checking the
Firmware
Version" in
"Setting"
"Checking and
Setting the Date
and Time".