Canon CXDI-40G COMPACT Service Manual page 128

Table of Contents

Advertisement

(2) Copy the "audit_grid.exe" file to the CCR folder of the Control Station.
(3) Start the CXDI host, and open the imaging screen.
(4) Perform the pre-imaging preparation and checks.
Adjust the distance between the focal point of the X-ray tube and the CXDI
sensor unit according to the grid being used.
No subject, full-screen exposure
Perform calibration and note that phototimer and other parts except for grid
that are located in the X-ray entrance surface are be installed in advance.
Do not perform calibration with the grid installed.
(5) Set the exposure dosage of the X-ray generator (same as calibration dosage).
(6) Press Patient Information ⇒ Imaging Part, and then press the X-ray Exposure button to
take three images.
(7) After imaging is completed, exit the CXDI host. (The three images are saved to dtstore.)
(8) Double-click the "audit_grid.exe" file in the CCR folder.
(9) The message "Input the File or Folder(#:Params,$:Initialize)?" is displayed. Drag and drop
onto the screen the first image file that was taken in step (6) and saved to dtstore.
* DICOM files must have the "dcm" extension.
* If a folder is dragged and dropped, all DICOM files in the folder become the target of
the study.
2. Installation
ƒ h ƒ ‰ ƒ b ƒ O • ƒ h ƒ ƒ b ƒ v
Drag and drop
- 108 -