8.5 The Connection with X-ray Generator Equipment

Photo coupler
RX_REQ
PC1
TLP281(GR/F)
Relay
RX_COM
RL1
NC2D-P-DC5V
Connections Between System and X-ray Generator
Connection conditions
1. The X-ray exposure signal line (including switching mechanism) must be insulated, and its
total impedance must be 100 ohms or less.
2. The maximum contact voltage of the X-ray exposure authorization signal line is AC 250V
and DC 30V, and its current ranges from 10mA to 2A.
Only the insulated secondary power supply can be connected.
3. The protective installation of the X-ray generator must have the same electrical potential as
the system.
2 Installation
POWER BOX
+5V(isolated)
Connector
J4
#2
#1
#4
#3
X-ray Generator
A2
YEL
A1
WHT
*1:Impedance between A1-A2 < 100ƒ ¶
B2
GRY
X-ray permission signal
*2:Permitted Vold AC250Vmax/DC30Vmax
Permitted Current 10mA-2A
B1
ORG
X-ray I/F Cable
- 38 -
X-ray request signal