9.3.4.2 Detailed Setup Procedures

Example: CXDI-40G COMPACT + CXDI-50G + Control PC
*1: Sensor1: CXDI-40G COMPACT
*2: Sensor2: CXDI-50G
*1: [Sensor1] section in screwcap.ini file
*2: [Sensor2] section in screwcap.ini file
Note: Do not connect a multiple number of sensor units simultaneously by following steps 1) to
3).
1)First, connect the first sensor unit (here, the CXDI-50G) to the control PC via the hub.
2) Change the IP address setting to "192.168.100.12" in order to allocate CXDI-50G as
[Sensor2]. The purpose of allocating the first sensor unit as [Sensor2] is to avoid conflict
between the two sensor IP addresses when the second sensor unit has been connected.
To change the setting, follow the instructions in the "Tool Software Operation Manual for
Ethernet" and then the instructions under "7. Imaging Unit IP Address setting."
3) Be absolutely sure to check that the IP address setting of the CXDI-50G has been changed.
To check the setting, use the ping command of the command prompt or follow the
instructions in the "Tool Software Operation Manual for Ethernet." and then the
instructions under "7. Imaging Unit IP Address setting."
4) Connect the second sensor unit (here, the CXDI-40G COMPACT) to the control PC via
the hub. There is no need to change the IP address of [Sensor1] since it is allocated to the
CXDI-40G COMPACT as the default.
5) After connecting the multiple number of sensor units, edit the SensorNum key in the
[SensorInfo] section of the screwcap.ini file to "SensorNum=2," and save the file.
6) After editing the screwcap.ini file, start up the system software.
7) Switch over to the CCR console menu on the DOS screen, select 1 => 0 => 7 to set "Max
Capture Devices=2," and input the fixed sensor serial numbers of the sensor units.
8) Upon completion of the CCR console menu setting, restart the system software.
This completes the connection of the multiple sensor units.
*Reference
When connecting three or four sensor units to the control PC, the units must be
connected one by one, and the above steps 1) to 8) must be followed to ensure that
there is no conflict between the sensor IP addresses by allocating the IP addresses in the
sequence of the [Sensor3] to [Sensor1] or [Sensor4] to [Sensor1] sections in the
screwcap.ini file.
If the number of CXDI-50G/50C imaging units is to be increased in an environment
where the control PC is connected to a single CXDI-40G COMPACT imaging unit,
they are connected with LAN cables via a hub. In a case like this, the additional sensor
units are first connected via the hub to the control PC, and their imaging unit IP address
settings are changed. After confirming the setting change, the existing sensor units must
be connected with LAN cables via the hub and the above steps 5) to 8) must be
followed.
2. Installation