Post-Installation Checks - Canon CXDI-40G COMPACT Service Manual

Table of Contents

Advertisement

15 Post-installation Checks

Check sheet
Checkpoint
Checking the imaging unit
Checking the date and time
Checking the software version
Identifying and registering the
imaging units
Inputting the control PC serial
number
Checking the operation unit
Checking the exposure
condition table
Checking the annotation
Network connections
Preparations prior to exposure
Checking image transfer to
printers and storages
Checking image transfer to
external memory device
Checking the image quality
2. Installation
Checkpoint details
Align the unit with the X-ray tube
Check the indicator mark for the X-ray tube
positioning.
Check that the unit does not interfere with the
cables.
Set the date.
Set the time.
Set the time zone.
Check that the CXDI application, firmware and
PLD code versions all match.
Register the serial numbers of the sensors
Input the serial number of the control PC to be used.
Set the contrast.
Set the brightness.
Set the gamma correction.
kV
mA
msec or mAs
Check that the setting have been made in
accordance with the customer's request.
IP address
Subnet mask
Default gateway
Perform calibration.
Perform self-test.
Check that the setting have been made in
accordance with the customer's request.
- Data checks
Checking that several sets of test data are transferred to
the storage and PACS and that there are no discrepancies
in the data
(1) The IDs, names and other data must be correct.
(2) The transfer sequence must be correct.
Check that the image is transferred properly.
Use SMPTE pattern to check the density on a linear
chart.
Check that there is no artifacts, shading, etc.
- 112 -
Check

Advertisement

Table of Contents
loading

Table of Contents