Canon CXDI-40G COMPACT Service Manual page 125

Table of Contents

Advertisement

4-6) When Setting Mode (0: Normal, 1: Expert) [0=0x0]: appears, select "0: Normal."
4-7) When CCR SETUP MENU appears, select "7 Scan Sensor Setup".
4-8) When Capture Device Configuration Table appears, enter the number of sensor to
which make the "Max Capture Device" recognized.
4-9) The dummies of "A/D Board Serial Number" are appeared: Enter the sensor serial
number for necessary type.
------A/D Board Serial Number 0 0 -> 50G : 0x199
------A/D Board Serial Number 0-1 -> 12 : 0x2009
------A/D Board Serial Number 0-2 -> 31 : 0x3002
------A/D Board Serial Number 0-3 -> 40G : 0x123
For example, when the upright stand model is connected to Sensor 1, and the 50G
cassette 14x17 connected to Sensor 2:
Enter "123" for "A/D Board Serial Number for Sensor ID#1"
Enter "199" for "A/D Board Serial Number for Sensor ID#2"
The CXDI application can now be launched on the control PC with the same conditions
in effect as if an imaging unit were connected.
@@@@@@@@ Capture Device Configuration Table @@@@@@@@
Max Capture Devices [2 = 0x2] : 2 <= Number of imaging units to be connected.
@@@@@@@@ Capture Device Configuration No.0 (SensorID#1 OPU)@@@@@@@@
------A/D Board Serial Number 0-0 -> 50G : 0x199
------A/D Board Serial Number 0-1 -> 12 : 0x2009
------A/D Board Serial Number 0-2 -> 31 : 0x3002
------A/D Board Serial Number 0-3 -> 40G : 0x123
A/D Board Serial Number for SensorID#1 [0x7900005 = 126877701] : 123
Custom Type[0:NO CUSTOM 1:STAND 2:TABLE 3:UNIV 4:CASSETTE 100um
5:CASSETTE 14X17_160um 6:CASSETTE 9X11_160um] [0 = 0x0] : 0
Field of View Rotation(0:No 1:Yes)
Constant for Exposure Index
@@@@@@@@ Capture Device Configuration No.1 (SensorID#2 OPU) @@@@@@@@
------A/D Board Serial Number 0-0 -> 50G : 0x199
------A/D Board Serial Number 0-1 -> 12 : 0x2009
------A/D Board Serial Number 0-2 -> 31 : 0x3002
------A/D Board Serial Number 0-3 -> 40G : 0x123
A/D Board Serial Number for SensorID#2 [0x10000004 = 268435460] : 199
Custom Type[0:NO CUSTOM 1:STAND 2:TABLE 3:UNIV 4:CASSETTE 100um
5:CASSETTE 14X17_160um 6:CASSETTE 9X11_160um] [0 = 0x0] : 0
Field of View Rotation(0:No 1:Yes)
Constant for Exposure Index
---- Need to re-start program to validate this change.