4. Connection to Units

4.1 CXDI-40G COMPACT Installation Flowchart
2. Installation
START
Affix the insulation sheets to
the imaging unit
Install the imaging unit
Install the phototimer
Install the power box and
control PC
Conduct calibration imaging
Install the grid
Conduct the self-diagnostic
test
END
- 11 -
* The Bucky unit (stand or table) and
imaging unit must be electrically isolated
from each other.
* Except for grid, phototimer and other
parts that are located in the X-ray entrance
surface must be installed when calibration
imaging is performed.
*Check the grid installation angle.