4-10) When CCR SETUP MENU appears, press the [Esc] key to return to Welcome to
CCR.
4-11) Select the command "8 - Exit" in the "Welcome to CCR" menu to exit the CXDI
application.
4-12) After exit the CXDI application and Windows XP desktop appears, starts the
command prompt screen (Start ⇒ Program ⇒ Command Prompt). And type the
command "ccrxxxxx /np" to start the CXDI application again.
a. When the CXDI application program is starting up, the message "There is no
BodyPart for SensorID#*. ** TYPE BodyPart will be created" appears. Click [OK].
(* differs according to the type of the sensor.)
The above message appears when a /np mode is used with a setting different to that
of the connected sensor.
b. The messages "System X-ray tube parameters (APERTURE,OFFSET) are wrong. will
you reset these wrong parameters??" and "The X-ray tube parameters
(APERTURE,OFFSET) in the BodyPart.ini file are wrong. Do you want to reset the
wrong parameters?" appear. Click [OK] for each.
These above messages might be appeared if /np mode is used with a different
setting.
5)
Going out of /np mode
Connect the sensor unit to the control PC.
Before using the system in normal condition, perform the following steps:
5-1) If /np mode has been used with a different setting as that of the connected sensor, as
mentioned in "(3) Notes", delete the BodyPart**.ini file made in the BodyParts folder,
and return the BodyPart**.ini file which has been moved into the BodyParts folder.
5-2) Enter the command "ccrstart.bat" on command prompt to boot the CXDI application.
Follow the procedure from 3) to 9) in previous section. And check these one will be
preformed properly without any problems.
5-3) Register the ccrstart.bat file to the StartUp.