9.3.2 Network Setting for Screwcap.ini
The CXDI software communicates with the imaging part through screwcap.dll by using the
communication protocol for sending and receiving commands and responses.
In conjunction with the communication, Screwcap.ini retains the information of the connected
imaging unit and control PC such as network addresses.
It is necessary to edit Screwcap.ini to communicate with the imaging part.
Screwcap.ini factory setting is the following default setting (see the figure below
sensor unit to be connected to the control PC, Screwcap.ini does not require editing, and only a
check need be conducted.
If a multiple number (2 or more) of sensor units are to be connected to the control PC,
Screwcap.ini must be edited. (* Refer to "9
sensor units.")
Screwcap.ini is located in the following directory:
D:\ccr\screwcap.ini
[HostInfo]
ResponseTimeout=30000
DataTimeout=60000
IntervalTimeout=30000
[SensorInfo]
SensorNum=1
[Sensor1]
IpAddress=192.168.100.11
CommandPort=12121
DataPort=12122
[Sensor2]
IpAddress=192.168.100.12
CommandPort=12121
DataPort=12122
[Sensor3]
IpAddress=192.168.100.13
CommandPort=12121
DataPort=12122
[Sensor4]
IpAddress=192.168.100.14
CommandPort=12121
DataPort=12122
[Log]
LogLevel=2
*1 "SensorNum=1" is the default value for the installer of the system software.
Only the [Sensor1] section is supported.
When the SensorNum key has been set to "2," "3" or "4" (SensorNum=2, 3 or 4), the
[Sensor1, Sensor2], [Sensor1, Sensor2, Sensor3] or [Sensor1, Sensor2, Sensor3, Sensor4]
sections are supported, respectively.
*2 Imaging unit IP address: This must be identical to the IP address stored in "9.3.3
Network settings stored in imaging unit" in the next section.
Note: All editing using keys other than the SensorNum keys is forbidden.
2. Installation
3.4. Connection setup for a multiple number of
.
[SensorInfo]
SensorNum=1*
In the case of a multiple
number of units, input here
the number of sensors to
be connected.
[Important]
If " 2" has been input for
SensorNum, the IP
address settings of the two
sensor units must match
the IpAddress key in either
the [Sensor1] or [Sensor2]
section without fail.
Imaging unit IP address.
The factory setting
This is identical to the IP address
written on the Di board inside the
sensor unit.
- 42 -
If a single
).
¹
*
²