Electromagnetic Compatibility

Guidelines and manufacturer's declaration - electromagnetic emissions
The concentrator is intended for use in the electromagnetic environment specified below.
The user of this device should make sure it is used in such an environment.
Emissions Testing
RF Emissions
CISPR11
RF Emissions
CISPR11
Harmonic emissions
CEI 61000-3-2 A
Voltage fluctuations /
flicker
IEC 61000-3-3 Compliant
Guidelines and manufacturer's declaration - electromagnetic immunity
The Horizon® P5 oxygen concentrator is intended for use in the electromagnetic environment
specified below. The customer or user of the HORIZON P5 oxygen concentrator should ensure
that it is used in such an environment.
Immunity Test
Electrostatic
± 8 kV in contact
Discharge (ESD)
± 15 kV in air
IEC 61000-4-2
DOMU722_v7_EN_20221006_Horizon® P5 User manual
Compliance Electromagnetic Environment - Guidelines
The Horizon® P5 oxygen concentrator uses RF
energy only for its internal functions. Therefore,
its RF emissions are very low and are not likely to
Group 1
cause interference in any nearby electronic
device.
Class B
The Horizon® P5 oxygen concentrator is suitable
for use in all premises, including domestic
A
premises and those directly connected to the
public
supplying domestic buildings.
Compliant
Test level
Compliance
IEC 60601
level
± 8 kV in
contact
± 15 kV in air
low-voltage
power
supply
Electromagnetic environment -
guidelines
The floors should be made of wood,
concrete or ceramic tiles. If the floors
are covered with synthetic materials,
the relative humidity should be at
least 30%.
network
41