When the power supply is disconnected from the AC outlet, also disconnect it from the
concentrator to avoid unnecessary discharge of the battery.
When charging a fully discharged battery, the charging process may start and stop for the
first few minutes. The fan will also stop.
4. Connect the nasal cannula to the cannula nozzle fitting.
The nozzle fitting is located on top of the Horizon® P5. The use of a single
inlet cannula up to 7.6 meters in length is recommended to ensure proper
breath detection and oxygen delivery. Additional titration may be required
to ensure proper oxygen delivery when using a particular cannula.
When using a 7.6 m (25 ft) long cannula with the Horizon® P5, an increase in flow setting
may be required.
Increasing the length of the nasal cannula may reduce the perceived noise during oxygen
bolus delivery.
To ensure oxygen flow, ensure that the nasal cannula is properly connected to the nozzle
fitting and that the tubing is not kinked or pinched in any way.
To confirm that air is being delivered to the end of the cannula, simply place your hand near
the air outlet at the end of the cannula; if you feel a puff, the air system is working properly.
If you do not feel gas flowing, check the cannula connections for leaks. Please contact your
supplier if this is the case.
To avoid choking and strangulation, keep tubes out of reach of children and pets.
Replace nasal cannulas regularly. Check with your supplier or physician to determine how
often the cannula should be replaced.