Horizon
®
P5 Column Replacement Procedure
Use only original columns. The use of unspecified columns may create a safety hazard
and/or adversely affect the performance of the device and will void your warranty.
The column replacement instructions should only be used when maintenance is required
and are not intended for training purposes. The column can be replaced by the patient by
following the procedure below.
Remove the current column (sieve beds)
1. Turn off the Horizon® P5 concentrator by pressing the "On/Off"
power button.
2. Remove the Horizon® P5 Concentrator from the carrying bag.
3. Remove the battery from the Horizon® P5 concentrator.
4. Turn the Horizon® P5 upside down, so that the display panel is facing
down and the bottom of the Horizon® P5 is facing up.
5. The column is the silver colored part (see picture on the right).
6. Remove the column by pressing the lock button with a finger.
7. While holding the button open, slide the column out by grasping the handle provided with
your other hand.
8. Remove the column completely from the Horizon® P5.
Installing the new column
9. Remove the dust caps from the new column. Make sure there is no dust or debris where the
dust caps were.
DOMU722_v7_EN_20221006_Horizon® P5 User manual