± 2 kV for power
Electrical fast
supply lines
transient / burst
± 1 kV for
EC 61000-4-4
input/output lines
± 1 kV phase to
Overvoltage
phase
transient
± 2 kV phase to
IEC 61000-4-5
earth
<0 % U
for 0.5 cycle
Voltage dips, short
interruptions and
0 % U
voltage variations
for 1 cycle
on power supply
input lines power
70 % U
supply lines
for 25 cycles
IEC 61000-4-11
<0% U
for 250/300 cycles
Power frequency
(50/60 Hz)
30 A/m
magnetic field
IEC 61000-4-8
NOTE: U
T
Guidelines and manufacturer's declaration - electromagnetic immunity
The Horizon® P5 oxygen concentrator is intended for use in the electromagnetic environment
specified below. The customer or user of the Horizon® P5 oxygen concentrator should ensure
that it is used in such an environment.
42
± 2 kV for
power
supply lines
± 1 kV for
input/output
lines
± 1 kV phase
to phase
± 2 kV phase
to earth
<0 % U
T
for 0.5 cycle
0 % U
T
for 1 cycle
70 % U
T
for 25 cycles
<0% U
T
for 250/300
cycles
30 A/m
- 50 or 60 Hz
- 50 or 60 Hz
is the a.c. main voltage prior to application of the test level.
The quality of the power supply
system should be that of a typical
commercial or hospital environment.
T
The quality of the power supply
should
be
commercial or hospital environment.
T
If the user of the HORIZON P5 oxygen
concentrator requires continuous
use during power outages, it is
T
recommended that the HORIZON P5
oxygen concentrator be powered
from an uninterruptible power
T
supply or battery.
The quality of the power supply
system should be that of a typical
commercial or hospital environment.
DOMU722_v7_EN_20221006_Horizon® P5 User manual
that
of
a
typical