Low Priority Alerts
The following low priority alerts are accompanied by a double beep every minute and a solid yellow
light.
Message
Condition / Action / Explanation
Low Battery
Charge Battery
Replace Column
Low oxygen
Warning
Hot battery
Battery
communication
error
Insufficient
ventilation
Cold system
DOMU722_v7_EN_20221006_Horizon® P5 User manual
Battery charge is low, between 5 and 10 %.
Connect an external power supply or turn off the power and insert a fully
charged battery.
Column (sieve beds) maintenance must be performed within 30 days.
Contact your service provider.
The concentrator produces oxygen at a slightly low level (between 82
and 87%) for 30 minutes. If the condition persists, contact your service
provider.
The temperature of the battery is above 40°C. Remove the battery from
the concentrator and allow it to cool in an open area for approximately
10-15 minutes. Then reinsert the battery into the device. If the problem
persists, contact your service provider.
The concentrator is producing oxygen but cannot indicate the battery's
state of charge. Check that the battery is connected to the device. If the
problem persists using the same battery, replace it with a new one, or
remove the battery and connect the concentrator to the external power
source.
The concentrator's temperature is above 50°C. Turn off the
concentrator, move it to a cooler room, and let it cool for 20 minutes. If
the alarm persists, check that the fan is running or call your service
provider.
The concentrator's temperature is below 0°C. Turn the device off, move
it to a warmer room, and let it warm up for 20 minutes before turning it
on. If the alarm persists, call your service provider.
29