Patient-to-patient disinfection
If used by multiple patients, the same cleaning routine must be performed between each patient,
and in addition, a complete disinfection procedure must be performed using a dry hydrogen peroxide
mist system and the Horizon® P5 disinfection protocol outlined in the Technical Manual. The
disinfection procedure is communicated to providers.
The cannula should be replaced routinely between patients.
Battery care
Your Horizon® P5 lithium-ion battery requires special care to ensure proper performance and long
life. Use only Horizon® P5 batteries with your Horizon® P5 concentrator.
Always keep the batteries dry with a 40-50% charge.
Always keep batteries away from liquids. If batteries become wet, discontinue use
immediately.
To extend the life of your battery, avoid using it in temperatures below 5˚C or above 40˚C
for extended periods of time.
When using multiple batteries, rotate them regularly
A battery kept alone without being used for more than 48 hours goes into a sleep mode
to limit its natural discharge. Plugging in the AC adapter allows the battery to wake up
immediately: you can then recharge the battery.
Your Horizon® P5 lithium-ion battery requires special care to ensure proper performance and long
life. Only use Horizon® P5 batteries with your Horizon® P5 concentrator.
Please follow these important guidelines to maximize battery performance and life:
Store the battery in a cool, dry place. Store it with a 40-50% charge.
If using multiple batteries, make sure each battery is labeled (1, 2, 3 or A, B, C, etc.) and
rotate them regularly.
Batteries should not be left on standby for more than 120 days at a time.
36
.