Recommended separation distances between portable and mobile RF communications
devices and the Horizon® P5 oxygen concentrator
The Horizon® P5 Oxygen Concentrator is intended for use in an electromagnetic environment in
which radiated RF interference is controlled.
The customer or user of the Horizon® P5 Oxygen Concentrator can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Horizon® P5 Oxygen Concentrator as
recommended below, based on the maximum transmit power of the communications
equipment.
Maximum rated
transmitter output
power
W
0,01
0,1
1
10
100
DOMU722_v7_EN_20221006_Horizon® P5 User manual
Separation distance according to the frequency of the transmitter
from 150 kHz to
80 MHz
d= 1,2√P
0,12
0,38
1,2
3,8
12
m
from 80MHz to
800 MHz
d= 1,2√P
0,12
0,38
1,2
3,8
12
from 800MHz to
2,5G Hz
d= 2,3√P
0,23
0,73
2,3
7,3
23
45