Scaleo medical HORIZON P5 User Manual page 32

Table of Contents

Advertisement

Message
Memory error
Sensor error
Fan error
Compressor
error
Troubleshooting
Solutions to some possible problems you may encounter are described in this section.
Problem
Any problem that is
associated with error
messages on the
concentrator, lights
and/or beeps.
The concentrator does
not turn on when the
on/off button is pressed.
32
Condition / Action / Explanation
Flash memory failed to read. Contact your service provider.
There is no communication with the oxygen sensor. Contact your service
provider.
The fan speed is different from the expected speed. Turn off the
concentrator and do not turn it on until it has been serviced by your
service provider.
The compressor is overspeeding or the pressure is above 32 PSI due to
the sieves beds. Turn off the concentrator and do not turn it on until it
has been checked by your service provider.
Possible Cause
Abnormal function
detected
The battery is
discharged or there is
no battery.
DOMU722_v7_EN_20221006_Horizon® P5 User manual
Recommended solution
See the "Notifications and Alerts" section.
Use an external power source or replace
the battery with a fully charged one.