ELECTRO MAGNETIC COMPATIBILITY (E.M.C)
• The POWEO® 150 requires special care regarding the E.M.C. it must be installed and commissioned by the
EMC information provided in this handout.
• The RF communication devices and portable phones may affect the lift POWEO® 150
• The use of accessories, transducers and cables other than those specified, except transducers and cables sold
by the manufacturer as replacement parts for internal components, can result in increased emissions or
decreased immunity of the device.
• It is appropriate that the patient lifter POWEO® 150 is not used next to other devices or stacked with them, if
it is not possible to do otherwise, the POWEO® 150 have to be monitored to verify normal operation in the
configuration in which it is used.
• In terms of the standard NF EN 60601-1-2: 2007, essential performance of the POWEO® 150 is to never
perform unwanted movements.
Directives and manufacturer's declaration - electromagnetic emissions
The POWEO 150 is designed to be used in the electromagnetic environment specified below. Should the client or
user of the POWEO 150 ensures it is used in such an environment
Emission test
RF Emissions
Group 1
CISPR 11
RF Emissions
Class B
CISPR 11
Harmonic
Not applicable
Emissions
IEC 61000-3-2
Voltage
Not applicable
fluctuations /
Flicker
IEC 61000-3-3
Conformity
The POWEO 150 uses RF energy only for its internal
functions. Therefore, its RF emissions are very low and are not
likely to cause interference in nearby electronic devices.
The POWEO 150 is suitable for use in all premises, including
domestic premises and those directly connected to the public
power network supplying domestic premises.
Electromagnetic environment - Guidelines