It is possible to deactivate the buzzer by briefly pressing the "Alarm" button. This locks the
buzzer on the screen and thus inhibits the alarm sound. The logo and the flashing LED will
still be active, and the error message will still be displayed on the screen.
It is always possible to change the setting when an alarm is active. The setting menu is displayed for
3 seconds, then the alarm message is displayed again. When the alarm indicating a change in the
column is active, a maintenance key flashes every 2 seconds at the top of the screen.
If a medium priority alarm occurs, it will shut down the device. Before the concentrator is completely
shut down, the corresponding error message is displayed for 10 seconds.
The following notification messages are accompanied by a single, short beep.
Message
Please wait
Shutdown in progress
28
Condition / Action / Explanation
The SCALEO Medical logo is displayed when the device is turned on.
The On/Off button has been pressed for two seconds and the system
starts up.
The On/Off button has been pressed for two seconds. The system will
shut down.